Development and Validation of a Chiral HPLC Method for Quantitative Analysis of Enantiomeric Escitalopram
Keywords:
Escitalopram, enantiomericpurity, method development, validation.Abstract
In the present study a rapid, accurate and precise chiral HPLC method was developed and validated for
enantiomeric separation of racemate citalopram and escitalopram according to the guidelines of United States of
Pharmacopeia (USP) and International Conference on Harmonization (ICH). The chiral chromatographic separation
was achieved with ammonium acetate/ ethanol/ 2-propanol/ methylene dichloride (100 : 150 : 70 : 30, v/v) at a flow
rate of 0.5 ml/min using a chiral CD-PH column. The HPLC analyses were monitored at 254 nm. The method showed
a good linearity with regression coefficient (r2) of 0.998 in the range of 20.0-70.0 μg/ml for escitalopram. The
detection limit (LOD), quantitation limit (LOQ) and average percentage of recovery for escitalopram were found to be
2.54, 7.68 μg/ml and 100.28% to 102.86%, respectively. The percentage of relative standard deviation (%RSD) for
intra- and inter- day precision were found as 0.16% and 0.09%, respectively. The established method proved as
reproducible with a %RSD value of less than 2 and having the robustness within specified limit. The present study
also showed the enantiomeric purity or excess (%ee) of seven pharmaceutical preparations of escitalopram. Thus the
proposed chiral method can be applied for the enantiomeric purity determination of escitalopram formulations.